Shipped May 11, 2021
Expected results: COV2105-1: Positive COV2105-2: Negative COV2105-3: Positive
Participant’s results
Reported | COV2105-1 | COV2105-2 | COV2105-3 |
---|---|---|---|
Positive | 21 (Acceptable) | 17 (Acceptable) | |
Negative | 12 (Unacceptable) | 34 (Acceptable) | 17 (Unacceptable) |
Inconclusive * | 2 (Unacceptable) | 1 (Unacceptable) | 1 (Unacceptable) |
Total | 35 | 35 | 35 |
Methods: 24 labs used Antigen detection methods – 11 labs used Nucleic Acid detection methods
CMPT QA/QC: Samples were prepared at CMPT; samples were tested for homogeneity and stability after the day of production. Samples were stable for 14 days at room temperature.