VR1908

VRE Screen: VR1908-1: Positive; VR1908-2: Positive; VR1908-3: Negative; VR1908-4: Positive  – Shipped: August12,2019

Participants’ results

ChallengePositiveNegativeIndeterminateNo reportTotal
VR1908-121 (Acceptable)3 (Unacceptable)1 (Unacceptable)25
VR1908-222 (Acceptable)2 (Unacceptable)1 (Unacceptable)25
VR1908-324 (Acceptable)1 (Unacceptable)25
VR1908-414 (Acceptable)7 (Unacceptable)2 (Unacceptable)1 (Unacceptable)25

CMPT QC/QA/Statistics: 

All samples were produced by CMPT. Samples were randomly selected for QC/QA. Statistics are not applicable in qualitative surveys. CMPT results are listed in the above table.

Samples VR1908 -1, and -2 (E.faecium) were van A positive, but van B negative. Sample VR1908-4 (E.faecium) was van B positive, but van A negative. Sample VR1908-3 (E.faecalis) was van A & B negative. All samples were homogeneous and stable for 20 days from the date of production, July 31, 2019.

Posted in critiques, Screening Bacteriology Results