TR1810

Trichomonas vaginalis – Shipped September 24, 2018

TR1810-1: Negative     TR1810-2: Positive    TR1810-3: Positive    TR1810-4: Negative

 

Participant’s results

ResultsTR1810-1TR1810-2TR1810-3TR1810-4
Positive34 (Acceptable)34 (Acceptable)
Negative34 (Acceptable)34 (Acceptable)
Total34343434

 

Suitability for grading

A challenge is considered suitable for grading if agreement is reached by 80 percent of selected reference group and at least 70 percent of the participants.

Trichomonas determination was correctly performed by 3/3 reference laboratories and greater than 70 percent of all laboratories and was thus, determined to be suitable for grading.


The samples are assessed for homogeneity and stability using in-house quality control methods and random selection of samples before and during production, and post sample delivery. The number of random samples selected is based on selection tables within Military standard 105E (1).

The challenge sample lot was confirmed to be homogeneous and stable for at least 15 days.

Grading of Trichomonas vaginalis is based on a Positive / Negative result, no further statistical analysis is performed on the results beyond that described under “Suitability for grading.”

(1) Famum NR. Acceptance sampling. In: Modern Statistical Quality Control and Improvement. Belmont, California.: Duxbury Press; 1994:305.

Posted in critiques, Trichomonas vaginalis Results